JWST Near Infrared Camera

The JWST Near Infrared Camera (NIRCam) offers imagingcoronagraphy, wide field slitless spectroscopy, and time-series monitoring both in imaging and spectroscopy, as well as wavefront sensing measurements for JWST mirror alignment.

On this page

The JWST Near Infrared Camera (NIRCam) observes from 0.6 to 5.0 μm and offers imagingcoronagraphy, wide field slitless spectroscopy, and time-series monitoring both in imaging and spectroscopy. NIRCam has 2 nearly identical modules pointing to adjacent fields of view. Each module uses a dichroic to observe simultaneously in a short wavelength channel (0.6–2.3 μm) and a long wavelength channel (2.4–5.0 μm).

 NIRCam has 5 observing modes for science:

NIRCam can also perform wavefront sensing measurements to align and phase JWST's primary mirror.

Figure 1. NIRCam field of view in the JWST focal plane

NIRCam in the JWST field of view



Observational capabilities

Each observing mode offers distinct capabilities and uses a subset of the optical elements available in the pupil and filter wheels.


Table 1. NIRCam observing mode parameters

Observing mode

Wavelength
coverage
(µm)

Field of view§

Pixel 
scale
(arcsec/pix)

Notes

Imaging

0.6–2.3

2 × 132" × 132"
(44" and 5" gaps)

0.031FWHM 2 pix
at 2.0 µm
2.4–5.0

2 × 129" × 129"
(48" gap) 

0.063FWHM 2 pix
at 4.0 µm

Coronagraphic
imaging

1.8–2.2 
2.8–5.0 

20" × 20"

0.031
0.063

Wide field
slitless spectroscopy
2.4–5.0 2 × 129" × 129"0.063R ~ 1,600 at 4 µm
Time-series imaging0.6–2.3 
2.4–5.0

129" × 129"
132" × 132"

0.031
0.063

Grism time series0.6–2.3
2.4–5.0
129" × 129"0.031
0.063

R ~ 300
R ~ 1,600 at 4 µm


§ Smaller fields of view are available using subarrays in the imaging, time series imaging, and grism time-series modes.

Field of view

Figure 2. NIRCam modules field of view

The 2 NIRCam modules image a 5.1' × 2.2' field of view with a central gap of 44"—the 2 adjacent 2.2' × 2.2' fields cover 9.7 arcmin² in total. The coronagraphs obtain images in regions of the sky outside the imaging/grism field of view. These are projected onto the detectors by optical wedges located on the pupil plane Lyot stopsCurrently, NIRCam coronagraphy is limited to Module A only.

Modules

NIRCam consists of 2 redundant modules—A and B—with identical optical elements and nearly identical throughput. The imaging and wide field slitless spectroscopy observing modes include options to use either a single module or both simultaneously with identical integration times, readout patterns, and filters. Currently, the coronagraphy and grism time-series modes are restricted to module A and time-series imaging is restricted to module B.

Channels

Each module includes a short (0.6–2.3 µm) and long (2.4–5.0 µm) wavelength channel. NIRCam uses a dichroic to observe at both wavelengths simultaneously in roughly the same field of view.


Table 2. Properties of short and long wavelength channels


Short wavelength channelsLong wavelength channels

Wavelength range

0.6–2.3 μm

2.4–5.0 μm

Fields of view

2 × 2.2' × 2.2' (with 4"–5” gaps)

2 × 2.2' × 2.2'

Pixels

8 × 2040 × 2040 pixels

2 × 2040 × 2040 pixels

Nominal pixel scale

0.031"/pixel

0.063″/pixel

PSF FWHM

0.07" @ 2μm (Nyquist)

0.13" @ 4 μm (Nyquist)

Grism slitless spectroscopy

R ~ 300

R ~ 1600 at 4 μm

Coronagraphic occulting masks

round: 2.1 μm

bar: 1.8–2.2 μm

round: 3.35, 4.3 μm

bar: 2.5–5.0 μm


Not including reference pixels insensitive to light that are 4 pixels along each outer edge.
PSF undersampled at lower wavelengths; slightly broader for coronagraphy.



Optical elements

Each channel includes a pupil wheel and a filter wheel, each with 12 optical elements, which may be used in various combinations. These elements include:

  • Filters with extra-wide, wide, medium and narrow passbands;

  • Coronagraphic Lyot stops to suppress light that passes the occulting masks;

  • Grism elements for wide field slitless and time-series spectroscopy in the long wavelength channel;

  • DHS element for time-series spectroscopy in the short wavelength channel;
  • Weak lenses to defocus light from bright sources to avoid detector saturation, also used for telescope alignment.

Within each module, light may (optionally) be passed through a coronagraphic occulting mask in the focal plane before being divided by the dichroic beam splitter to the short and long wavelength channels.

Figure 3. NIRCam pupil and filter wheels

The NIRCam pupil and filter wheels contain a total of 48 optical elements. Filters are color-coded by wavelength. Wider filters in the figure appear more transparent than narrower filters.

Filters

The 29 NIRCam filters have names "Fnnnx" where "nnn" refers to the central wavelength (e.g., 220 for 2.20 µm) and "x" refers to the filter width: "W2" for extra-wide (R ~ 1), "W" for wide (R ~ 4), "M" for medium (R ~ 10) and "N" for narrowband (R ~ 100).

Figure 4. NIRCam and JWST optical telescope element (OTE) filter throughputs

Click on the figure for a larger view.

Total system throughput for each NIRCam filter, including contributions from the JWST Optical Telescope Element (OTE), NIRCam optical train, dichroics, filters, and detector quantum efficiency (QE). Throughput refers to photon-to-electron conversion efficiency. Averages across all detectors are plotted. The vertical gray bar marks the approximate dichroic cutoff between the short and long wavelength channels. Filters marked "P" are located in the pupil wheel, requiring transmission through a second filter in the filter wheel, either F150W2, F322W2, or F444W. In these cases, the combined transmissions are plotted. (Figure version 6.0: May 16, 2024)



Detectors

NIRCam has 10 Teledyne HgCdTe H2RG detectors, or sensor chip assemblies (SCAs), each with 2040 × 2040 pixels sensitive to light. Eight short wavelength detectors cover roughly the same area on the sky as the 2 long-wavelength detectors. In full frame imaging mode, all 10 detectors are read out non-destructively every 10.74 s. The smallest supported science subarray (64 × 64 pixels) can be read out in 49 ms (the shortest exposure time).



Sensitivity

In a ~10 ks image, NIRCam reaches S/N ~ 10 on point sources as faint as ~8 nJy (AB mag 29.1) in some wide filters. See NIRCam Imaging Sensitivity.

Please use the Exposure Time Calculator (ETC) for the most up-to-date sensitivity estimates for specific proposed observations. See the NIRCam Sensitivity article for more information.

Figure 5. NIRCam sensitivity for imaging

Sensitivity is shown as S/N = 10 detection limits for point sources in a ~10 ks image (comprised of 10 exposures, ~1 ks each). The sources are assumed to have flat spectra in nJy (and AB magnitudes). Zodiacal light is assumed to be 1.2 times the minimum. Filter widths are shown as horizontal bars. Extra-wide/wide, medium, and narrow filters are labeled in normal, bold, and italic text, respectively each with progressively thicker bars. Please use the Exposure Time Calculator (ETC) to calculate sensitivity estimates for your specific proposed observations.


Saturation

In standard imaging mode NIRCam bright source saturation limits range between ~9 and ~16 Vega magnitudes, depending on the filter. Subarrays allow for shorter readout times to reach brighter magnitudes. Brighter magnitudes can also be reached in the time-series observing modes, as weak lenses (at short wavelengths) and grisms (at long wavelengths) spread the bright light over more detector pixels. Stars observable without saturation (80% full well) include those visible to the unaided eye: ~5th magnitude or brighter, depending on the observing mode, wavelength, and configuration.

Please use the ETC to calculate saturation estimates for your proposed observations.



Data calibration and analysis

Information about calibration is available in the JWST Science Data Overview article. This article and its links point to content about absolute astrometricflux, and wavelength calibration, as well as information on calibration reference files.

Details about the structure and format of the data can be found in the Understanding JWST Data Files article. The JWST calibration pipeline is described in both the JWST Science Calibration Pipeline article and the official software documentation. More information about post-pipeline processing can be found in the JWST Post-Pipeline Data Analysis article.



External NIRCam links and documents

STScI NIRCam website

University of Arizona NIRCam website

NASA NIRCam site

JWST Technical reports  

Lectures

JWST Community Lecture Series - NIRCam: Your Next Near-Infrared Camera (M. Rieke)



Acknowledgements

NIRCam was built by a team at the University of Arizona (UofA) and Lockheed Martin's Advanced Technology Center, led by Prof. Marcia Rieke at UoA.



References

Rigby, J. et al. 2023, PASP, 135, 048001
The Science Performance of JWST as Characterized in Commissioning

Rieke, M. et al. 2023, PASP, 135, 028001
Performance of NIRCam on JWST in Flight

Rieke, M. et al. 2005, SPIE, 5904, 1
Overview of James Webb Space Telescope and NIRCam's Role

Rieke, M. et al. 2003, SPIE, 4850, 478
NGST NIRCam Scientific Program and Design Concept

Beichman, C. et al. 2012, SPIE, 8442, 2
Science opportunities with the near-IR camera (NIRCam) on the James Webb Space Telescope (JWST)

NIRCam Papers from the August 2005 SPIE meeting




Notable updates
  •  
    Added information on using the DHS for grism time series

  •   
    Merged this page with NIRCam Overview

  •  
    Updated grism resolution

  •  
    Updated pixel scale values.
Originally published